    氧化鋯氧量分析儀使用說明書(英文版)

    來源:安徽天分儀表有限公司   2017年06月24日 09:12  

     

    Check whether the packaging is intact

     
      

    According to the packing list to check the number of instruments and accessories are complete

    Ø  If the instrument is found to have been damaged during transport, the carrier and supplier should be notified immediay. Do not damage the instrument back to the consignor, otherwise the carrier will not be compensated. Keep the box and damaged parts for carrier inspection.

     
      

    The instrument has been accuray calibrated before shipment and no further calibration before installation.

     
      

    Failure to follow the operating instructions for the use of an oxygen analyzer may result in unnecessary danger.

     
      

    Before making electrical connections, make sure that all wires are insulated and that the power cord and signal cable are connected in strict accordance with the operating instructions to avoid electric shock.

     Ø The protective ground on the power plug must be well grounded to ensure operator safety. Oxygen analyzer before the first start, the ground must be checked to ensure that the wire without interruption.

     
      

    Special protective measures must be taken when operating on high pressure vessels. When working on high temperature containers and pipes, remember to wear protective masks to protect your face and eyes from harm.

     
      

    In the vicinity of high temperature containers and pipes must wear protective clothing, wearing protective gloves, to ensure safety.

     
      

     

     

     

    Standard

     

    ISO 3864, No. B.3.6

     

     

     

    Standard

     

    IEC 417, No. 5017

    Marking instructions

    Be careful of electric shock

    Marking instruct

      Ground point

     

     

     

    Standard

     

    ISO 3864, No. B.3.1

     

     

     

    Standard

     

    IEC 417, No. 5041

    Marking instructions

        attention

    Marking instruct

    Pay attention to high temperature

     

     

     
     

     

     

    Oxygen analyzer for boiler, kiln flue gas oxygen content (concentration) of continuous detection, the system uses non-sampling (wet) analysis.

     The integrated zirconia oxygen analyzer consists of a detection unit (oxygen probe) and a control unit (transmitter). Wherein the detection unit is installed on the flue and is equipped with a temperature control unit and a zirconia sensor; the control unit is a terminal for the user to operate the integrated function, including an operation panel, a display interface, and a 4-20mA current output port.

    The probe part is made of stainless steel, durable, standard configuration length of 1.0m, according to user needs to provide a length of 0.2m ~ 2.0m probe.

    The analyzer is not properly serviced and is wired correctly according to the wiring diagram. Analyzer installation is simple and convenient, installed in the high temperature flue work, should pay attention to take the appropriate security measures.       

     

     

     

     

     

     

     

      Oxygen analyzer

     

    Split oxygen analyzer is divided into two parts: analysis and detection unit and control unit.

     

     
     
     
     

    Control unit (transmitter) shell for the cast aluminum, the shell has a control circuit.

     

     

     

     

     

                       Control unit specification

    The control unit should be installed in locations where vibration and high temperatures should be avoided. Strong vibration and high temperatures can have a significant impact on the life of the instrument.

     

     

     
     

     

     

    TFZO-5/05YT-GD  xygen Analyzer Analysis Unit The probe section is made of stainless steel, the head shell is cast aluminum, and the housing has an electrical connection terminal. In addition, the external reference air and calibration gas inlet and exhaust are also located on the cast aluminum housing.

      Oxygen analyzer detection units with standard lengths of 0.5m, 1.0m and 1.2m are available. Other specifications can be customized production.

     

     

     

     

     

     

     

                Analysis of the detection unit specifications

    Ø

     

       1The flue gas temperature of the test unit installation point should meet the relevant requirements. Generally speaking, the flue gas temperature is low, the detector has long service life, high flue gas temperature and short service life. Flue gas temperature is best at 300-600 ° C.

     

       2The detection unit can not be installed in the flue gas does not flow dead, can not be installed in the flue gas flow quickly (such as some bypass airway expansion chamber).

     

       3Requiring the flue leakage is small, the detector installation and maintenance convenience, for small and medium-sized boilers, it is recommended to install the preheater in the economizer, because the boiler system flue gas flow from the furnace to the drum, after the superheater, The air preheater is discharged from the chimney by the draft fan after recycling. If the measuring point is too close to the flue gas furnace outlet, because the temperature is too high, the flow rate is quick, the detector stainless steel shell will form erosion erosion, shorten the service life; if the measuring point is too partial, due to flue system leakage phenomenon, Will result in high oxygen value at the measuring point, can not accuray reflect the furnace oxygen in the flue gas.

     

     

    The oxygen analyzer can be mounted via a flange, including

     

     
     

     

     

     

    Equipment installation flange

     

     

     

     

     

        Equipment installation flange specifications

    Oxygen analyzer welding equipment has been installed flange, aperture 4 × Φ14mm, screw hole spacing of 130mm.

    Ø Adapter flange mounting short connection

     

     

     

     

     

     

     

     

                    Adapter flange mounting shorting

    Made in the form of a fitted flange, it is suitable for flue with flanged flanges on flue.

     

     

     

     

                           Wearable casing specifications

     

     

     

     

    The use of corrosion-resistant casing flange installation, suitable for all the structure of the flue, aperture 4 × Φ14mm, screw hole pitch of 130mm.

     

     

     

     

                     

                        Use sleeve flange to install

     

     

     

     

     

     

     

     

     

     

     

     

     

    [1] In the flue to open a diameter and casing diameter of the hole.

        [2]Place the casing on the flue and leave enough space to insert and tighten the safety bolts.

        [3]Install the flange gasket and insert the oxygen analyzer.

        [4]Insert the safety bolts from the rear end of the sleeve flange into the sleeve flange, flange pad and device mounting flange, and install the upper and lower nuts and nuts.

     

    Ø Install the flange and the weld is coated with anti-corrosion protective layer, can not be removed. The installation will be different from the installation steps, such as flue with a welding flange, the user will need to bring their own parts.

     

     

     

     

     

       

    Range: 0.01 ~ 25.0% O2 (programmable)

    Display: LCD

    Language: Chinese / English (for English menu, please specify when ordering)

    Output signal: 4 ~ 20mA

         Probe material: 316L

    Protection class: IP65

    Repeatability: ± 0.5% of full scale

    Basic error:  ± 1% (full scale)

    Stability:  ± 1% (instrument continuous test 4h)

    Response time: access to standard gas T  3s to reach 90% of the response

    Ambient temperature: -10  ~ 65

    Power supply: 220V AC 50 ~ 60Hz

    Power: 150W

    Sample gas temperature:  700

    Sample gas pressure range: ± 20KPa

    Self-diagnosis information: thermocouple reverse, thermocouple split even, heating abnormalities, the upper and lower temperature alarm, the upper and lower limits of oxygen alarm.

     

     

     
     

      

     

       Zirconia is a high-temperature electrolyte concentration battery, in the hundreds of degrees of high temperature environment, can produce oxygen ion transport conductivity, due to the measured gas (flue gas or other gas) and the reference gas (air or other gas) In the zirconia on both sides of the platinum electrode oxygen partial pressure is different between the two electrodes have a certain amount of oxygen ions migrate and produce oxygen concentration potential, the potential value and the relationship between oxygen concentration, can be Nernst (Nernst) formula Said:

                 E=RT/4F×LnP1/P2

     WhereE—Oxygen concentration potentialV

          R—Ideal gas constant8.314J/moLK

          T—Absolute temperature valueK

          F—Faraday constant96500c/moL

          P1—Reference gas partial pressureair

    P2—Measured gas partial pressure

     

     

     
     

     

     

     

     

     

     

     

     

    Cable selection:

    Integrated zirconia oxygen analyzer using RVV 4 × 1.5 two-core cable, of which: two 220V power supply, the other two then 4-20ma current output.

     

     
     

     

     

    After the correct installation is complete, turn on the power, until the normal temperature, about 20 minutes later, the temperature constant temperature 700 . You can normally analyze the oxygen content, this time the oxygen content is the oxygen content of flue gas.

               When the new equipment is shipped from the factory, it has been debugged and checked. After installation, it is not necessary to do any calibration and debugging. When the equipment is used for a certain period of time, if the calibration is necessary, the following method can be used.

     

     

     

     

     
     

     

     

     

    (Air flow  100ml / min), or the oxygen probe from the furnace directly placed in the air, until the data is stable, press the "SET" key to enter the main menu, press the "" Press the "  " button to enter the calibration submenu, press the "" or "" key to select the "Auto Calibration" menu, press the  key to  "" After entering the background automatic calibration, press“”  key to return to the measurement state. Oxygen shows about 20.58% can be. Then screw the seal screw on the standard gas inlet.

     

    Ø This step must be performed automatically every time the

       detection unit or the control unit is replaced.

     

     

     

     

     
     

     

     

    Press the "SET" key to enter the main menu, press the "" key to enter the main menu, press the "" key to enter the main menu, press the "" Or "" key to select the "Oxygen parameter correction" menu, press "" key to enter the calibration submenu, press "" or "" to select the "Standard Auto Calibration" menu, press "" key to enter To enter the calibration page, press "" or "" to correct the data to the same standard value. Press "" key to finish the calibration. Press "  " key to return to the measurement state. Then screw the seal screw on the standard gas inlet.

     

     

     

     

    Press "SET" key to enter the main menuPress "" or "" to select the "Oxygen Upper Limit Alarm" menu, press "" key to enter the adjustment page, press "" or "" to adjust the oxygen quantity Upper limit data, press "" key to complete the adjustment, press "   " key to return to the measurement state.

     

     

     

     

     

    Press "SET" key to enter the main menuPress "" or "" to select the "Oxygen Lower Limit Alarm" menu, press "" key to enter the adjustment page, press "" or "" to adjust the oxygen content Lower limit data, press "" key to complete the adjustment, press "   " key to return to the measurement state.

     

     

     

     

    Press "SET" key to enter the main menuPress the "" or "" key to select the "temperature upper limit alarm value" menu, press "" key to enter the adjustment page, press "" or "" key to adjust the temperature upper limit data, press "" key to complete the adjustment, "    " Key to return to the measurement state.

     

     

     

    Press "SET" key to enter the main menuPress "" or "" to select the "lower temperature alarm value" menu, press "" key to enter the adjustment page, press "" or "" key to adjust the temperature lower limit data, press "" key to finish the adjustment, "   " Key to return to the measurement state.

     

     

    Press "SET" key to enter the main menuPress "" or "" to select the "Oxygen Range Setting" menu, press "" key to enter the adjustment page, press "" or "" to adjust the oxygen range data (range from 5% -25% Press the "" key to complete the adjustment, press the "   " key to return to the measurement state .4-20mA current output will automatically and range corresponding.

     

     

    Ø When the instrument is factory, the range is set to 0-25% by default.

     

     

    Press "SET" key to enter the main menuPress "" or "" to select the "temperature setting" menu, press "" key to enter the adjustment page, press "" or "" key to adjust the constant temperature (constant temperature point 700  and 750  two Select), press "" key to complete, press "    " key to return to the measurement state.

    Ø When the instrument is shipped, the constant temperature

         is set to 700 ° C by default

     

     

     

    Press "SET" key to enter the main menuPress the "" or "" key to select the "Temperature alarm control" menu, press "" key to enter the adjustment page, Press "" or "" to select the alarm "ON" or "OFF", Press "" key to complete,press "   " to return to the measurement state.

     

     

    Press "SET" key to enter the main menuPress "" or "" to select the "Oxygen Alarm Control" menu, press "" key to enter the adjustment page, press "" or "" to select the alarm "ON" or "OFF",Press "" key to complete,Press "    " key to return to the measurement state.

     

     
     

     

     

    Zirconia analyzer in the warehouse storage, should be packaged intact, storage temperature of 0 ~ 40 , relative humidity of not more than 85%, no strong corrosive gases.

     

     
     

     

     

    Range 20.6%

    Range 25.0%

    Range 20.6%

    Range 25.0%

    Oxygen content

    Current mA

    Oxygen content

    Current mA

    Oxygen content

    Current mA

    Oxygen content

    Current

    mA

    0.1

    4.08

    0.1

    4.06

    9

    10.99

    9

    9.76

    1

    4.78

    1

    4.64

    9.5

    11.38

    9.5

    10.08

    2

    5.55

    1.5

    4.96

    10

    11.77

    10

    10.40

    2

    5.55

    2

    5.28

    11

    12.54

    11

    11.04

    2.5

    12.37

    2.5

    12.62

    12

    13.32

    12

    11.68

    3

    6.33

    3

    5.92

    13

    14.10

    13

    12.32

    3.5

    6.72

    3.5

    6.24

    14

    14.87

    14

    12.96

    4

    12.54

    4

    13.30

    15

    15.65

    15

    13.60

    4.5

    7.50

    4.5

    6.88

    16

    16.43

    16

    14.24

    5

    7.88

    5

    7.20

    17

    17.20

    17

    14.88

    5.5

    8.27

    5.5

    7.52

    18

    17.98

    18

    15.52

    6

    8.66

    6

    7.84

    19

    18.76

    19

    16.16

    6.5

    9.05

    6.5

    8.16

    20

    19.53

    20

    16.80

    7

    9.44

    7

    8.48

    20.6

    20.00

    20.6

    17.18

    8

    10.21

    8

    9.12

    22

     

    22

    18.08

    8.5

    10.60

    8.5

    9.44

    23

     

    23

    18.72

    9

    10.99

    9

    9.76

    24

     

    24

    19.36

     

     

     

     

    25

     

    25

    20

     

     

    temperature()

    Oxygen content%

    Oxygen potential E(mv)

    temperature()

    Oxygen content%

    Oxygen potential E(mv)

    700

    0.5

    77.94

    750

    0.5

    81.95

    700

    1

    63.41

    750

    1

    66.67

    700

    2

    48.88

    750

    2

    51.39

    700

    3

    40.38

    750

    3

    42.46

    700

    3.5

    37.15

    750

    3.5

    39.06

    700

    4

    34.35

    750

    4

    36.12

    700

    4.5

    31.89

    750

    4.5

    33.52

    700

    5

    29.68

    750

    5

    31.20

    700

    5.5

    27.68

    750

    5.5

    29.10

    700

    6

    25.86

    750

    6

    27.18

    700

    6.5

    24.18

    750

    6.5

    25.42

    700

    7

    22.62

    750

    7

    23.79

    700

    7.5

    21.18

    750

    7.5

    22.27

    700

    8

    19.83

    750

    8

    20.84

    700

    8.5

    18.55

    750

    8.5

    19.51

    700

    9

    17.36

    750

    9

    18.25

    700

    9.5

    16.22

    750

    9.5

    17.06

    700

    10

    15.15

    750

    10

    15.93

    700

    10.5

    14.13

    750

    10.5

    14.85

    700

    11

    13.15

    750

    11

    13.83

    700

    11.5

    12.22

    750

    11.5

    12.85

    700

    12

    11.33

    750

    12

    11.91

    700

    12.5

    10.47

    750

    12.5

    11.01

    700

    13

    9.65

    750

    13

    10.14

    700

    13.5

    8.86

    750

    13.5

    9.31

    700

    14

    8.10

    750

    14

    8.51

    700

    14.5

    7.36

    750

    14.5

    7.74

    700

    15

    6.65

    750

    15

    6.99

    700

    15.5

    5.96

    750

    15.5

    6.27

    700

    16

    5.30

    750

    16

    5.57

    700

    16.5

    4.65

    750

    16.5

    4.89

    700

    17

    4.03

    750

    17

    4.23

    700

    17.5

    3.42

    750

    17.5

    3.59

    700

    18

    2.83

    750

    18

    2.97

    700

    18.5

    2.25

    750

    18.5

    2.37

    700

    19

    1.69

    750

    19

    1.78

    700

    19.5

    1.15

    750

    19.5

    1.21

    700

    20.6

    0.00

    750

    20.6

    0.00


     

    安徽天分儀表有限公司

    Anhui TianFen Instrument Co., Ltd.

    電子:2229496671

    :2229496671

    +86-

    :+86-

    地址:安徽省天長市經(jīng)濟(jì)開發(fā)區(qū)經(jīng)三路*創(chuàng)業(yè)服務(wù)中心

    Add:Tianchang City, Anhui Province Economic Development Zone by the three high-tech business service center

     

     

    免責(zé)聲明

    • 凡本網(wǎng)注明“來源:儀表網(wǎng)”的所有作品,均為浙江興旺寶明通網(wǎng)絡(luò)有限公司-儀表網(wǎng)合法擁有版權(quán)或有權(quán)使用的作品,未經(jīng)本網(wǎng)授權(quán)不得轉(zhuǎn)載、摘編或利用其它方式使用上述作品。已經(jīng)本網(wǎng)授權(quán)使用作品的,應(yīng)在授權(quán)范圍內(nèi)使用,并注明“來源:儀表網(wǎng)”。違反上述聲明者,本網(wǎng)將追究其相關(guān)法律責(zé)任。
    • 本網(wǎng)轉(zhuǎn)載并注明自其它來源(非儀表網(wǎng))的作品,目的在于傳遞更多信息,并不代表本網(wǎng)贊同其觀點(diǎn)或和對其真實(shí)性負(fù)責(zé),不承擔(dān)此類作品侵權(quán)行為的直接責(zé)任及連帶責(zé)任。其他媒體、網(wǎng)站或個(gè)人從本網(wǎng)轉(zhuǎn)載時(shí),必須保留本網(wǎng)注明的作品第一來源,并自負(fù)版權(quán)等法律責(zé)任。
    • 如涉及作品內(nèi)容、版權(quán)等問題,請?jiān)谧髌钒l(fā)表之日起一周內(nèi)與本網(wǎng)聯(lián)系,否則視為放棄相關(guān)權(quán)利。
    聯(lián)系我們

    客服熱線: 15267989561

    加盟熱線: 15267989561

    媒體合作: 0571-87759945

    投訴熱線: 0571-87759942

    關(guān)注我們
    • 下載儀表站APP

    • Ybzhan手機(jī)版

    • Ybzhan公眾號

    • Ybzhan小程序

    企業(yè)未開通此功能
    詳詢客服 : 0571-87759942
    主站蜘蛛池模板: 国产美女主播一级成人毛片| 天天影院成人免费观看| 久久亚洲精品成人综合| 国产v片成人影院在线观看| 亚洲精品亚洲人成人网| 99精品国产成人一区二区| 成人无遮挡裸免费视频在线观看| 成人国产一区二区三区精品| 国产成人av一区二区三区在线观看| 亚洲欧洲精品成人久久曰影片 | 欧美成人亚洲高清在线观看| 成人免费看www网址入口| 亚洲激情成人网| 成人亚洲欧美日韩在线观看| 亚洲av无码成人网站在线观看| 成人毛片一区二区| 久久久久免费看成人影片| 成人性生交大片免费看| 91成人精品视频| 亚洲欧美日韩成人高清在线一区| 成人性生话视频| 成人综合激情另类小说| 美国特级成人毛片| 亚洲欧美成人综合| 国产成人精品一区二区三在线观看| 欧美成人精品第一区二区三区| 免看**一片成人123| 国产v片成人影院在线观看| 国产精品成人久久久久久久| 欧美成人午夜免费完成| 精品久久久久成人码免费动漫| 国产成人8X视频网站入口| 国产成人精品A视频一区| 国产成人啪精品视频免费网| 国内外成人免费视频| 国产成人十八黄网片| 成人av在线一区二区三区| 国产成人精品三级麻豆| 亚洲欧洲精品成人久久曰| 亚洲AV成人无码天堂| 综合558欧美成人永久网站|