SH-IVD
- Wide range (0 – 10ppm to 0 – , Selection of 2 ranges out of 18 ranges)
- 2 patterns of portable and process
- LED digital display with versatile display functions
- Equipped with self-diagnosis function
- Simple, one-push calibration
Use
- Air separation plants
- Ferrite incinerators
- Oxygen enriching equipment
- Gas displacement monitoring
- Gas purity control (H2, He, Ar, ...)
- Heat treatment of iron and steel and other metals (non-oxidation furnaces)
- Special metal welding atmosphere gas (welding process)
- Semiconductor process atmosphere gas
- Inactive gas for seals
Specifications
Model | SH-IVD |
---|---|
External dimensions | 202(W)×165(H)×342(D) |
Panel cut | 190(W)×150(H) |
Painting color | Case: Munsell N3 semigloss, Panel: Munsell 5Y7/1 semigloss |
Range | Selection of 2 ranges out of 16 ranges (0 – 1, 2, 5, 10, 25, O2) (0-10, 20, 50, 100, 200, 500, 1,000, 2,000, 5,000ppm) |
Output | 0 – 10mV or 0 – 1V, 4 – 20mA (Isolate linear output) |
Sample gas flow rate | 500ml/min |
Sample gas pressure | 3 – 100kPaG (Sample gas flow rate at 500ml/min) Pump required at ≦3kPaG, Pressure reducing valve required at ≧100 kPaG |
Sample gas conditions | No inclusion of combustible gas or corrosive gas, No condensation at ambient temperature |
Reproducibility precision | % range: ±0.5%FS, ppm range: ±1%FS |
Response time | % range: Within 5 sec (90% response), ppm range: Within 10 sec (90% response) |
Ambient temperature | -5 to 40°C |
Power voltage | 100VAC±10%, 50/60Hz, 200VA |
Weight | Approx. 7kg |
TF-SD
Sensor Model AS-210
Receiver Model DTF-201
- Separation enabled between the sensor and the receiver
- Stand-alone panel type available with the combination of sampling materials
- No need of standards gas or reference air (air 1-point calibration)
- Reduction zone measurement type also available
Use
Measurement of oxygen concentration in oxidized combustion exhaust gas
- Small- and medium-sized boilers
- Small- and medium-sized industrial heating furnaces
- Coke ovens
- Heat-treating furnaces (radiant tube furnaces)
Measurement of atmosphere concentration in the reduced combustion exhaust gas
- Ceramic kilns (shuttle furnaces)
- Ceramic kilns (tunnel furnaces)
- Ceramic raw material firing furnaces (rotary kilns)
Specifications
Model | TF-SD | |
---|---|---|
Measurement principle | Zirconia 2-cell form | |
Measurement range | 0 – 5/10/25% , -15/0/25%, -6/10%, -3/5% | |
Output | 1st output | Selection from among 4 – 20mADC |
2nd output | Selection from among 0-1 VDC / 0-5 VDC / 0-10 VDC | |
Response time | Within 10 sec (90% response at calibration gas changeover) | |
Alarm display | Content display on the receiver with Error No. | |
Warming-up time | 3 min | |
Power source | 100V (85 – 132V) AC, 50/60Hz, Approx. 50VA | |
External dimensions | 96(W)×96(H)×170(D)(Receiver) |